Fogging Tester

The Fogging Tester is specially designed equipment for testing the fogging characteristics or VOCs of leather, plastics, vinyl and textiles for automotive, aerospace and other transportation vehicles.This fogging tester comprises of following equipment.

OTHER DETAIL
  • Thermostatically Controlled Bath: Available with all safety devices for over- heating, this is designed to operate at up to 130°C. All the parameters like bath capacity, circulation system and heating system work in coordination as to control the temperature within + 0.5°C throughout the bath. A slow and uniform speed is maintaining for the agitation of the bath.
  • Note: There is a centrifugal pump at the bottom of some of the heaters and circulators that enable the pumping of liquid at high speed around the bath. Passing liquid in the beakers have different speeds and this will cause differentiation in the temperatures of different beakers. The distance between the bath fluid and lower surface of the glass plate should be 60 + 2 mm.
  • Cooling Plates: Hollow in structure, these cooling plates are made from rust-proof metals with aluminum on the side facing the glass plate. These are specially designed for keeping the glass plates cool. The cooling water flows through a whole of the plate with two connections. A minimum weight of 1kg should be maintained in the cooling plate filled with to overcome the buoyancy of the beaker in the bath. For each different beaker, a separate cooling plate is used. Designing of the cooling plates and water thermostat is done in such a way that the mean temperature of water remains at 21°C and the temperature difference between the inlet and outlet does not exceed 1°C.
  • Flat bottomed beakers: These are made from heat-resistant glass
  • Metal rings: Made from rust proof steel, these rings are used to keep the test pieces flat.
  • Sealing Rings: These rings are L- shaped or circular, made from silicone or fluoro rubber.
  • Float-glass plates: These plates are available in residential or windshield window quality for condensation of the fogging. Available in the tin and not- tin finish, these glass plates have reflectometer values within + 2% units.
  • Note: Identification of the tin and non-tin surfaces of the glass plates can be done by viewing in a darkened room under a UV light at 254 nm wavelength. When exposed to UV light, the tin surface of the plate will fluoresce. Filter paper
  • Aluminum foil discs: These discs are stored in a dedicator for avoiding condensation and contamination
  • Reflectometer: also known as gloss meter
  • Spacer: This type of spacer is specially installed during reflectometer measurements to prevent contact with the condensate on the glass plate. These are made from high-grade materials such as paper or plastic with a circular hole for the measurements.
  • Dishwasher
  • Balance
  • Polyethylene gloves
  • Desiccators: These are available with suitable drying material
TECHNICAL SPECIFICATIONS
  • External diameter: 80mm
  • Internal diameter: 74mm
  • Height: 10mm
  • Mass: 55+1g
SEALING RINGS
  • Inner diameter: 90mm to 95mm
  • Thickness: 2mm to 4mm
  • Hardness: 50 IRHD to 70 IRHD
FLOAT-GLASS PLATES
  • Thickness: (3 + 0.2) mm
  • Square with minimum dimensions: (110×100) mm
  • Circular with diameter: 103mm
FILTER PAPER
  • Diameter: 110mm
  • Mass per unit surface: 90 g/m2
Scroll to Top